Comparison Chart Details

This comparison chart shows each generators capabilities, specifications, and uncertainty values to help you determine which suits your calibration needs best.
Note: on the Model 2500, the “ST” model is not shown, the differences in this model are the chamber size which is 15" x 15" x 12" (38.1 cm x 38.1 cm x 30.5 cm) plus on this model you can add a high air flow (HAF) of 10 to 40 Liters per minute.

Generator Comparison Chart

SPECIFICATIONSModel 1200Model 2500Model 2900Model 9500Model 3920
Relative Humidity Range:10% to 95%10% to 95%10% to 98%5% to 98%0.00003% to ~50%
Frost Point Temperature Range:–18 to 0 °CRequires external PC and software-33.60 to 0 °C-32 to 0 °C-95 to 0.01 °C
Dew Point Temperature Range:–20 to 50 °CRequires external PC and software-36.94 to 70 °C-35 to 70 °C-50 to 10 °C
Temperature Range:10 to 60 °C0 to 70 °C-10 to 72 °C0 to 72 °C-80 to 12 °C
Temperature Control Stability:±0.04 °C0.01 °C0.002 °C0.01 °C0.008 °C
Temperature Uniformity:0.1 °C0.1 °C when operating at chamber temperatures within 10 °C of abmient0.04 °C0.04 °CN/A
Temperature Heating Rate:2 Minutes per °C (average)2.5 Minute per °C (average)2 Minute per °C (average)0.5 °C per Minute (average)2 Minute per °C (average)
Temperature Cooling Rate: (to 0 °C)4 Minutes per °C (average)2.5 Minute per °C (average)2 Minute per °C (average)0.5 °C per Minute (average)2 Minute per °C (average)
Temperature Specification:0.05 °C0.06 °C0.027 °C0.018 °C0.027 °C
Gas Type:Air or NitrogenAir or NitrogenAir or NitrogenAir or NitrogenCO2 free Air or Nitrogen with ambient pressure frost point <–80 °C
Gas Pressure Rating: (MAWP)174 psiG175 psiG175 psiG350 psiG80 to 300 psiG
Gas Flow Rate Range:2 to 10 L/min5 to 20 L/min10 to 50 L/min10 to 100 L/min0.5 to 10 L/min
Gas Flow Rate Specification:1.0 L/min2% of full Scale5% of full scale2% of full Scale2% of full scale
Saturation Pressure Range:Ambient to 152 psiAAmbient to 150 psiAAmbient to 160 psiAAmbient to 325 psiA~Ambient to 250 psiA
Saturation Pressure Specification:0.08% of FS psiA±0.075 psiA (low range)±0.032 psiA±0.0045 psiA (low range)0.05 psiA
±0.225 psiA (high range)±0.03 psiA (high range)
Test Chamber Pressure Range:AmbientAmbientAmbientAmbient to 20 psiAAmbient to 50 psiA
Test Chamber Pressure Specification:0.08% of FS psiA±0.075 psiA0.02% of reading±0.0023 psiA0.02% of reading
Display Resolution:0.010.010.0010.0010.001
Test Chamber Dimensions:6" x 6" x 6" (152 mm x 152 mm x 152 mm)12" x 12" x 10" (305 mm x 305 mm x 254 mm)12" x 12" x 10" (305 mm x 305 mm x 254 mm)12" x 12" x 12" (305 mm x 305 mm x 305 mm)N/A
Physical Dimensions:11.85" H x 24.10" W x 14.11" D (30.1 cm x 61.2 cm x 35.8 cm)19” H x 33” W x 20” D (482.6 mm x 838.2 mm x 508 mm)22" H x 36" W x 23" D (558.8 mm x 432 mm x 584.2 mm)38.25” H x 60” W x 36” D (971.5 mm x 1524.0 mm x 914.4 mm)37.9" H x 23" W x 34.9" D (965mm x 585mm x 889mm)

UNCERTAINTY
SPECIFICATIONSModel 1200Model 2500Model 2900Model 9500Model 3920
Dew/Frost Point Uncertainty:0.1 °CRequires external PC and software0.08 °C/0.05 °C0.05 °C(Dew –50 to 10 °C) 0.01% |R| + 0.07 °C
(Frost >= –90 °C) 0.05% |R| + 0.07 °C
(Frost <–90 °C) 1.7% |R| –1.415 °C
RH Uncertainty:0.5%0.5%0.5% of reading0.3% of readingN/A
Test Chamber Temperature Uncertainty:0.05 °C0.06 °C0.031 °C0.021 °C0.034 °C


Note: The specifications listed, and the information contained within this web page are subject to change without notice.